18
InfiniiMax III+ Series Probes User’s Guide
1
Using InfiniiMax III+ Series Probes
Both types, when used together, provide a significant level of ESD protection. Of
the two, only the table-mat and wrist-strap combination provides adequate ESD
protection when used alone. To ensure user safety, the static-safe accessories
must provide at least 1 M
W
of isolation from ground. Purchase acceptable ESD
accessories from your local supplier.
Safely connecting the probe to an oscilloscope
To protect against ESD damage, always use the four steps shown in
page 19 when connecting your probe to the oscilloscope.
Probing the DUT
When making your measurements, you’ll often need to probe different locations on
the DUT. You can safely move any of the following three probe heads
without
having to first break the amplifier-to-head connection:
• N5445A differential browser head
• N5439A ZIF head,
• N5444A 2.92 mm/3.5 mm/SMA head.
• N2848A QuickTip InfiniiMode head.
The only exception is when the DUT is
not
grounded to the oscilloscope via the AC
mains ground. In this case, connect the DUT ground to the oscilloscope ground
before
moving the probe. An example of a device having a floating ground would
be a battery-powered DUT.
WARNING
These techniques for a static-safe work station should not be used when working on
circuitry with a voltage potential greater than 500 volts.
CAUTION
When connecting a probe head to a probe amplifier, push straight in. When
disconnecting a probe head from an amplifier, pull the probe head connectors straight
out of the sockets. Never bend the probe head in order to pry it loose from the
amplifier. Also, do not wiggle the probe head up and down or twist it to remove the
connectors from the sockets. This can damage the pins in the amplifier or the probe
head itself.
CAUTION
When probing with an N2836A and N5441A, always disconnect the probe head from
the amplifier
before
unsoldering, moving to a new position, and resoldering the head.
This is because some soldering-iron tips can hold a charge which can damage the
probe amplifier.
CAUTION
When probing with an N2848A, always disconnect the probe head from probe tip
before
unsoldering, moving to a new position, and resoldering the tip. This is because
some soldering-iron tips can hold a charge which can damage the probe amplifier.
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...