InfiniiMax III+ Series Probes User’s Guide
1
Using InfiniiMax III+ Series Probes
To Avoid Damaging the Probe
In this section, you’ll learn to properly handle your probes to prevent damage and
maintain high performance. For more safe-handling information, go to
http://www.keysight.com/find/esd-best-practices-demo
Using a static-safe work station
InfiniiMax probes and accessories are ESD sensitive devices and should be treated
with care. Before using or handling the probe or accessories, always wear a
grounded ESD wrist strap and ensure that cables and probe heads are discharged
before being connected.
All work, including connecting probe amplifiers to the oscilloscope, should be
performed at a static-safe work station as shown in
.
CAUTION
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy
high-performance probes, resulting in costly repairs. Always wear a wrist strap when
handling probe components.
CAUTION
Probes are sensitive devices and should be treated with care. Do not bend or kink the
probe amplifier cable. Do not drop heavy objects on the probe, drop the probe from
large heights, spill liquids on the probe, etc. Any of these examples can significantly
degrade the performance of the probe.
CAUTION
When storing the probe, it is best to coil the cable in a large radius and avoid a net
twist in the cable during the process. This can be done in a similar manner to how
garden hoses or extension cords are typically coiled.
CAUTION
InfiniiMax I and II probe heads cannot be used with Infiniimax III+ probe amplifiers and
InfiniiMax III+ probe heads cannot be used with InfiniiMax I and II amplifiers.
CAUTION
Never allow the probe head to be connected to the probe amplifer, if the probe
amplifier is
not
connected to the oscilloscope channel.
CAUTION
Always disconnect an N2836A or N5441A solder-in probe head from the probe
amplifier before unsoldering, moving to a new position, and resoldering the head.
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...