– Setting Change: delay timer start at the end of a settings change in voltage, current, or output state. Press
OCP
Start Set
.
Clears and OVP and OCP Event
To clear the protection function, first remove that condition that caused the protection fault.
Press
Protection Clear
to clears the protection function and returns the output to its previous operating state.
From the remote interface:
To set the over-voltage protection for output 1 to the maximum limit:
VOLT:PROT MAX (@1)
To enable the over-current protection for output 1 and 3:
CURR:PROT:STAT ON (@1,3)
To set the over-current protection delay time for output 1 at 3 seconds:
CURR:PROT:DEL 3 (@1)
To set the over-current protection delay timer start for output 1 to CC transition:
CURR:PROT:DEL:STAR CCTR (@1)
To clear protection for output 1:
OUTP:PROT:CLE (@1)
Keysight E36300 Series User's Guide
61
Содержание E36300 Series
Страница 1: ...User s Guide E36300 Series Programmable DC Power Supplies ...
Страница 12: ...12 Keysight E36300 Series User s Guide ...
Страница 29: ...Front panel Rear panel E36312A and E36313A only Keysight E36300 Series User s Guide 29 ...
Страница 106: ...106 Keysight E36300 Series User s Guide ...