Programming Ranges
The following table shows the maximum voltage, current, and power that can be programmed for each model.
Parameter
+6V output
+25V output
-25V output
1
Output Identifier
P6V
P25V
N25V
Output Number
1
2
3
Voltage
MAXimum
6.18 V
25.75 V
0 V
MINimum
0 V
0 V
-25.75 V
DEFault (*RST)
0 V
0 V
0 V
Current
MAXimum
5.15 A
10.3 A
2
1.03 A
2.06 A
2
1.03 A
2.06 A
2
MINimum
0.001 A
0.001 A
0.001 A
DEFault (*RST)
5 A
10 A
2
1 A
2 A
2
1 A
2 A
2
Note:
1. Applicable for E36311A or when SCPI ID is set to E3631A (E36312A and E36313A). In normal operation, the
E36312A and E36313A will output as +25 V.
2. Applicable for E36313A.