5
Triggering and Gating
140
Keysight 53210A User’s Guide
Querying the Gate Time
The gate time set by the
expected value
and
resolution
parameters can be
determined after execution of the
CONFigure
or
MEASure
command.
For example, assume a measurement requiring 9-digit (10 mHz) resolution of a 5
MHz signal is required. Because digits of resolution is
approximately
equal to
Log
10
(
expected value
) -
Log
10
(
resolution
), this measurement could be configured
as:
MEAS:FREQ? 5e6, 5E-3, (@1)
// expected freq, resolution
A typical measurement using this configuration might return:
+4.99999949006
With the counter display registering:
4.999 999 50MHz
(9 digits)
Querying the gate time after sending this command returns:
SENS:FREQ:GATE:TIME?
+1.00000000000000E-001
(100 ms)
From
this corresponds to 9 digits of resolution. To achieve a desired
resolution (digits) in terms of
gate time
, locate the number of digits in
and select the corresponding gate time:
SENS:FREQ:GATE:TIME 100e-3 // set gate time = 100 ms
SENS:FREQ:GATE:SOUR TIME
// set gate source
As another example, assume a measurement requiring 7-digit resolution of a 5 ns
(200 MHz) signal. Recalling that
Log
10
(
expected value
) -
Log
10
(
resolution
) is
approximately equal to digits of resolution, the measurement can be configured
as:
MEAS:PER? 5E-9, 5E-16, (@1) // per meas of 200 MHz signal
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...