Formats and Data Flow
7
Keysight 53210A User’s Guide
209
Press the ‘
Select
’ soft key to select the directory or folder.
3
Enter the file name. The knob scrolls through A-Z (upper case), a-z (lower
case), numbers 0-9, selected keyboard characters, decimal point (.),
underscore (_), and space. Once the desired character is set, press the right
arrow (>) key under the knob to move to the next position.
4
Repeat until the file name is complete. Use the left arrow key (<) to backspace
and change a character. Changing the character to a space erases the
character.
5
Press ‘
STORE STATE
’ to store the state under the file name defined. A .sta
extension is added to the file name.
6
States (state files) are recalled by pressing the ‘
RECALL STATE
’ soft key,
highlighting the file name, and pressing ‘
Select
’. The state is recalled
immediately.
Storing User Preferences
Non-volatile I/O settings and user preferences can be stored and recalled using
the commands:
MMEMory:STORe:PREFerences <file>
MMEMory:LOAD:PREFerences <file>
<file>
is created during execution of the
STORe
command. The parameter format
is “[
drive:path]<file_name>
”.
drive
is either
INT
(internal flash memory) or
USB
(external memory device).
path
is an absolute path and folder name. If a folder is
specified, it must have been previously created (see
MMEMory:MDIRectory
).
The combination of folder and file name cannot exceed 240 characters, and
cannot contain the characters \ / : * ? “ < > |. Preference files have a .prf extension.
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...