7
Formats and Data Flow
196
Keysight 53210A User’s Guide
2
all readings are
stored
in an internal counter format - the programmed format
(
FORMat
subsystem) is set when readings are transferred to the output buffer.
3
if
reading
memory
overflows the first (oldest) readings are overwritten and the
‘Reading Mem Ovfl’ bit (14) in the Questionable Data Register is set. The most
recent measurements are retained.
Output Buffer
Data is available to the PC over the LAN, USB, and GPIB interfaces once it is in the
output buffer (
). The following section explains how readings are
transferred from memory to the buffer. If during a transfer the
output buffer
fills,
command execution is paused until readings are removed by the PC - creating
room in the buffer. No readings are lost.
Transferring Readings from Memory to the Output Buffer
The following commands transfer readings from volatile reading memory to the
buffer.
READ?
- is equivalent to executing
INITiate:IMMediate
directly followed by
FETCh?
. With
READ?
, once
all
readings specified by
TRIGger:COUNt
and
SAMPle:COUNt
are complete and in memory, they are instantaneously read into the
output buffer
(
If the data format is REAL (
FORMat
subsystem), each block of readings generated
by
READ?
is preceded by an IEEE 488.2 Indefinite-Length block header (
).
As such, when sending multiple commands in a single command string,
READ?
should always be the last command in the string.
READ? Example
// frequency meas of 500 kHz signal with uHz resolution
CONF:FREQ 500E3, 1E-6, (@1) // configure counter
TRIG:COUN 1
// set trigger count
SAMP:COUN 10
// take 10 readings
READ?
// read readings when complete
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...