5
Triggering and Gating
142
Keysight 53210A User’s Guide
External Gate Signal Polarity
When using any external gate source the polarity of the gate signal and therefore,
the duration of the gate, is set or changed with the command:
[SENSe:]FREQuency:GATE:POLarity {POSitive|NEGative}
[SENSe:]FREQuency:GATE:POLarity?
(query form)
POSitive
starts the measurement on a positive edge on the Gate In/Out BNC or
channel 1 input, and stops the measurement on the
next
negative edge.
NEGative
starts the measurement on a negative edge on the Gate In/Out BNC or
channel 1 input, and stops the measurement on the
next
positive edge.
CONFigure
and
MEASure
do not change the polarity setting. Following a reset
(
*RST
) or counter preset (
SYSTem:PRESet
), the negative edge is selected.
To set the polarity of the external gate signal:
// select the gate polarity and source
CONF:PER
SENS:FREQ:GATE:POL POS // set polarity
SENS:FREQ:GATE:SOUR EXT // set source
NOTE
OUTput:STATe
OFF
must be set when using gate source
EXTernal
Gate Signals on the ‘Gate In/Out’ BNC
” for more information and for information
on using gate signals to synchronize other instruments.
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...