6
53210A Math, Graphing, and Data Logging
158
Keysight 53210A User’s Guide
CALCulate[1]:AVERage:STATe {OFF|ON}
CALCulate[1]:AVERage[:STATe]?
(query form)
ON
- enables statistical calculations on the readings as they occur. Statistical data
includes: mean, standard deviation, maximum value, minimum value, and average
peak-to-peak value.
OFF
- disables statistics computation.
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
), statistical
computations are disabled.
Reading Count
Statistical calculations start when
Math
and the
Statistics
function are enabled.
The number of readings on which a given set of statistics are based is shown as
the
Count
value (
). The count can also be queried with the command:
CALCulate[1]:AVERage:COUNt:CURRent?
The count can be read at any time after the measurements have been initiated (i.e.
INITiate:IMMediate
,
READ?
,
MEASure?
).
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
), the count is set to
‘1’.
Mean, Standard Deviation, Minimum Value, Maximum Value
The mean (average), standard deviation, minimum value, and maximum value for
the current reading count can be determined with the command:
CALCulate[1]:AVERage:ALL?
If the readings have been scaled (see “
”), the statistics are based
on the scaled readings.
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...