111
Chapter 6 Disassembly and Repair
Self-Test Procedures
4
6
626 - 629
626: Self-test failed; waveform filter path select relay
627: Self-test failed; -10 dB attenuator path
628: Self-test failed; -20 dB attenuator path
629: Self-test failed; +20 dB amplifier path
These errors indicate that the specified relay is not being properly
switched or the attenuator/amplifier is not providing the expected
amplification or gain. These self-tests use the internal ADC to verify that
attenuators are operating properly.
630
Self-test failed; internal ADC over-range condition
This error indicates a probable ADC failure. The failure could be of the
system ADC, the ADC input multiplexer, or the ADC input buffer
amplifier.
631
Self-test failed; internal ADC measurement error
This error indicates a probable ADC failure. The failure could be of the
system ADC, the ADC input multiplexer, or the ADC input buffer
amplifier.
632
Self-test failed; square/pulse DAC test failure
This error indicates a probable failure of the square/pulse DAC.
Содержание 33210A
Страница 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 10: ...8 ...
Страница 13: ...1 1 Specifications ...
Страница 19: ...2 2 Quick Start ...
Страница 37: ...3 3 Front Panel Menu Operation ...
Страница 52: ...50 3 ...
Страница 53: ...4 4 Calibration Procedures ...
Страница 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Страница 97: ...5 5 Block Diagram ...
Страница 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Страница 103: ...6 6 Disassembly and Repair ...
Страница 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Страница 129: ...7 7 Backdating ...