116
Chapter 6 Disassembly and Repair
Mechanical Disassembly
6
Removing the Main Power Supply Assembly
Loosen the captive screw securing the power supply cover to the deck.
Slide the power supply cover and power supply and lift from the deck.
Slightly spread the ends of the power supply cover and slide the power
supply out of the cover. Disconnect the line input, ground, and output
cables from the power supply. The main power supply should be replaced
as an assembly.
WARNING
Always be sure to re-attach the green ground wire to the power supply
before operating the instrument.
Captive screw