78
Chapter 4 Calibration Procedures
Internal ADC Adjustment/Self Calibration
4
Internal ADC Adjustment/Self Calibration
The function generator stores calibration constants related to the gain
and offset of the internal ADC. Setup 6
must
always be performed
before
any later adjustments are attempted. The internal ADC is then
used as a source for the calibration constants generated in setup 7.
1
Make the connections as shown below.
2
Set the DMM to display 5 1/2 digits.
3
Enter the following setup, and then measure the dc value.
* Constants are stored after completing this setup.
4
Enter the measured value of the dc source using the numeric keypad or
knob.
Note
This setup requires approximately 15 seconds to complete.
5
Disconnect all cables from the rear panel Modulation In connector.
Nominal Signal
Setup
DC level
6*
~1.1 Vdc ±10%
Calibrates the internal ADC.
Modulation In