43
Chapter 3 Front-Panel Menu Operation
To Unsecure and Secure for Calibration
4
3
To Secure After Calibration
1 Select the Secure Code interface.
Press
and then select the
Test/Cal
softkey.
2 Enter a Secure Code.
Enter up to 12 alphanumeric characters. The first character must be
a letter.
Use the knob to change the displayed character. Use the arrow keys to
move to the next character.
3 Secure the Instrument.
Select the
Secure
softkey.
4 Exit the menu.
Press the
DONE
softkey.
Use the knob to change the displayed character.
and
Use the arrow keys to move to the next character