Chapter 4 Calibration Procedures
Automating Calibration Procedures
4
4
Automating Calibration Procedures
You can automate the complete verification and adjustment procedures
outlined in this chapter if you have access to programmable test
equipment. You can program the instrument configurations specified
for each test over the remote interface. You can then enter read-back
verification data into a test program and compare the results to the
appropriate test limit values.
You can also adjust the instrument from the remote interface. Remote
adjustment is similar to the local front-panel procedure. You can use a
computer to perform the adjustment by first selecting the required
function and range. The calibration value is sent to the instrument and
then the calibration is initiated over the remote interface. The instrument
must be unsecured prior to initiating the calibration procedure.
For further information on programming the instrument, see chapters
3 and 4 in the
Keysight
33210A User’s Guide.
Содержание 33210A
Страница 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 10: ...8 ...
Страница 13: ...1 1 Specifications ...
Страница 19: ...2 2 Quick Start ...
Страница 37: ...3 3 Front Panel Menu Operation ...
Страница 52: ...50 3 ...
Страница 53: ...4 4 Calibration Procedures ...
Страница 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Страница 97: ...5 5 Block Diagram ...
Страница 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Страница 103: ...6 6 Disassembly and Repair ...
Страница 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Страница 129: ...7 7 Backdating ...