81
Chapter 4 Calibration Procedures
AC Amplitude (high-impedance) Adjustment
4
4
AC Amplitude (high-impedance) Adjustment
The function generator stores a calibration constant for each high-
impedance attenuator path. The gain coefficient of each path is
calculated using two measurements; one with the waveform DAC at +
output and one with waveform DAC at – output. The setups, therefore,
must be performed in pairs.
1
Connect the DMM as shown below.