![Keysight Technologies U3022AM06 Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380084.webp)
U3022AM06
Troubleshooting the Test Set
RF Switching Failures
If the Test Set’s RF signal functionality is not operating properly, use the following
procedure to aid in isolating the source of problem. The internal RF switches
used in this instrument are biased with the +24 Vdc supply voltage and controlled
logic signals coming from the Test Set Controller board and Switch Interface
board.
1.
Place the network analyzer and Test Set side by side (not stacked) and connect
the I/O cable (N4011-21002). See
for I/O cable picture.
2.
Ensure that the network analyzer is
not
in Multiport Mode.
3.
Using the Test Set I/O Data command values to be given in the preceding
procedure, you will observe S21 response measurements.
RF Switching Path Test
If you suspect an RF signal path problem with the Test Set and have verified that the
problem in not the front panel RF interface cables, the following procedure will check
all of the RF signal paths through the Test Set.
•
Install the Test Set I/O cable from the analyzer to the Test Set’s rear panel.
•
Remove the front panel RF interface cables and reinstall the analyzer’s front panel
jumper cables.
Equipment Required
Gather the equipment and accessories listed below:
•
PNA, PNA-L or PNA-X Network Analyzer
•
Two RF Flex Cables (3.5 mm male)
•
3.5 mm Adapters (female to female)
•
Test Set I/O Cable
84
Keysight U3022AM06 User's and Service Guide
Содержание U3022AM06
Страница 6: ...Table of Contents...
Страница 7: ...U3022AM06 Keysight U3022AM06 User s and Service Guide 7...
Страница 80: ...Figure 66 U3022AM06 System Block Diagram U3022AM06 Option 129 80 Keysight U3022AM06 User s and Service Guide...
Страница 101: ...Keysight U3022AM06 User s and Service Guide 101...