U3022AM06
Controlling the Test Set
Interface Control Mode
Interface Control mode will not function properly when using multiport mode.
Multiport mode will reset the switch path commands of the interface control. It is
recommended that the network analyzer be restarted in stand-alone mode if the
interface control is being used.
Overview of the Interface control
The Interface Control feature allows you to send data and remote commands to
control external equipment using the GPIB, Handler I/O, Test Set I/O, Dwell After
Command and Aux I/O without the need of an external computer. Refer to analyzer’s
Help menu, “Rear Panel Tour” and “Interface Control.”
•
A unique set of control data can be sent for each channel. In addition, a unique
set of control data can be sent before the channel sweep starts and after the
sweep ends.
•
Interface Control settings can be saved and recalled from the “Interface Control”
dialog box or with Instrument State Save and Recall.
•
Interface Control settings can be copied to other channels using Copy Channels.
•
Control data can only be WRITTEN to the interfaces, NOT READ from the
interfaces.
•
Control data is sent in the following order and this order cannot be changed:
Refer to the Help menu.
1. GPIB Commands - BEFORE
2. Handler I/O Control
3. Test Set I/O Control (addr.data)
4. Dwell After Command
5. Aux I/O Output Voltage
Keysight U3022AM06 User's and Service Guide
53