![Keysight Technologies U3022AM06 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380055.webp)
U3022AM06
Controlling the Test Set
Using Interface Control Mode
An Instrument Preset will reset all of the fields to their default settings.
If an error is encountered when sending Interface Control data, an error message is
displayed on the analyzer’s screen and the Channel Trigger State is set to Hold. You
must fix the condition that caused the error, then change the Channel Trigger State
to its original setting.
Figure 48
Interface Control
Enable Interface
Control: (1)
Test Set I/O Control
(addr.data): (2)
Enables and disables ALL Interface Control communication. When cleared (default
setting) Interface Control is disabled and NO data is sent. To send data, the
individual interfaces must also be enabled.
Provides control of the Test Set I/O Interface on the rear panel of the analyzer. Used
to control your Test Set.
Only positive integers are allowed to select switch positions or states of DUT control
interface lines. Refer to
“Address and Data Values” on page
Address and data integers must be separated by a period. Each set of entries should
be separated by a new line, or carriage return. The front panel Enter key can be used
to insert a new line into the field. The quantity of Test Set I/O entries that can be
entered is limited by the available memory of the analyzer.
Address and Data example: addr.data
(3)
0.0
16.1
32.2
(1)
(5)
(4)
(6)
(2)
(8)
(3)
(7)
(8)
Keysight U3022AM06 User's and Service Guide
55
Содержание U3022AM06
Страница 6: ...Table of Contents...
Страница 7: ...U3022AM06 Keysight U3022AM06 User s and Service Guide 7...
Страница 80: ...Figure 66 U3022AM06 System Block Diagram U3022AM06 Option 129 80 Keysight U3022AM06 User s and Service Guide...
Страница 101: ...Keysight U3022AM06 User s and Service Guide 101...