U3022AM06
Cal Kit Operational Check
Figure
61
Typical Reflection Tracking with E8362B/C or N5230A/C
Figure 6
2
Typical Reflection Tracking with N5242A/B (Ports 1 - 2)
Response from 10 MHz to 500 MHz is normal due to the PNA or PNA-X couplers in
comparison to the Test Set bridges. The bridges have more gain in the coupled RF
path.