U3022AM06
Installation Procedures, Test Set to Analyzer
illustrates the final two digits of the part number for each cable. The cables
must be connected in the numeric order listed in
.
Kit U3021-60047 contains eight extra cables that are not listed above and are not
used in this interface connection
Figure 26
N5242A/B RF Interface Cable Connections
Table 11
N5222/42A/B RF Interface Cable Connection (U3021-60046 or U3021-60047)
Numeric
Order
RF Cables
From: PNA or PNA-X
To: Test Set
1
U3042-20031
Port 1 SOURCE OUT
Ports 3, 5 and 7 SOURCE IN
2
U3042-20032
Port 1 CPLR THRU
Port 3, 5 and 7 CPLR THRU
3
U3042-20029
Port 1 CPLR ARM
Port 3, 5 and 7 CPLR ARM
4
U3042-20030
Port 1 RCVR A IN
Port 3, 5 and 7 RCVR OUT
5
U3042-20041
Port 2 CPLR ARM
Ports 4, 6 and 8 CPLR ARM
6
U3042-20043
Port 2 SOURCE OUT
Port 3, 5 and 7 SOURCE IN
7
U3042-20044
Port 2 CPLR THRU
Port 3, 5 and 7 CPLR THRU
8
U3042-20042
Port 2 RCVR B IN
Port 3, 5 and 7 RCVR OUT
29 30 32 Port 9
Port 10 44 42 41
30 32
42 44
31
31
29
43
41
43
36
Keysight U3022AM06 User's and Service Guide