General Specifications
Typical Reflection Tracking
Specifications for the Test Set are typical. System performance for the Analyzer and
Test Set are only characteristic and intended as non-warranted information.
Reflection Tracking is one of the Error Terms available after performing a Test Port
"1-Calibration." It is derived from the following measurement parameters:
Insertion Loss, Source and Receiver Path Loss, and Network Analyzer Mixer
Calibration.
Typical performance is based on 1 to 2 units performance. Refer to
Table 7
Table 8
Frequency
dB
300 kHz to 10 MHz
1
1.
N5230A/C only.
−0.2
10 MHz to 4 GHz
−3
4 GHz to 6 GHz
−4
6 GHz to 10.5 GHz
−5
10.5 GHz to 13.5 GHz
−6
13.5 GHz to 15 GHz
−7
15 GHz to 20 GHz
−10
Typical Reflection Tracking for the E8362B/C or N5230A/C with the Test Set
Typical Reflection Tracking for the N5242A/B with the Test Set
Frequency
dB
10 MHz to 50 MHz
−0.5
50 MHz to 500 MHz
−2
500 MHz to 3.2 GHz
−3
3.2 GHz to 10 GHz
−5
10 GHz to 16 GHz
−7
16 GHz to 20 GHz
−10
20 GHz to 24 GHz
−11
24 GHz to 26.5 GHz
−12
18
Keysight U3022AM06 User's and Service Guide
Содержание U3022AM06
Страница 6: ...Table of Contents...
Страница 7: ...U3022AM06 Keysight U3022AM06 User s and Service Guide 7...
Страница 80: ...Figure 66 U3022AM06 System Block Diagram U3022AM06 Option 129 80 Keysight U3022AM06 User s and Service Guide...
Страница 101: ...Keysight U3022AM06 User s and Service Guide 101...