![Keysight Technologies U3022AM06 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380035.webp)
U3022AM06
Installation Procedures, Test Set to Analyzer
RF Interface Cable Connection (U3021-60046 or U3021-60047)
RF Interface Cable Connections
illustrates the cable configuration of the Test Set to a 4-Port
analyzer. The cables have been supplied with the kit (U3021-60046 or 60047).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8)
on the PNA or PNA-X. The RCVR R1 to SOURCE OUT reference loop jumper (x2 or
x4) remain on the front panel.
2.
Connect the RF interconnect cables from the PNA or PNA-X to the Test Set in the
order listed. As you are connecting each cable, torque to 8 in-lb. The longer,
straight end of each cable is connected to the Test Set. Refer to
and
3.
Add the front panel port labels from the cable kit (U3021-60046) over the
Analyzer's port numbers.
•
Port 9 over Port 3 (U3022-80006)
•
Port 10 over Port 4 (U3022-80008)
Over torque will cause damage to the Test Set and may cause connectors to spin or
become loose.