Installation Note E8356-90001 7
In This Guide
Operating Precautions
Figure 1
ESD Protection Setup
The following table lists Keysight part numbers for ESD protection devices. To
order, refer to
“Contacting Keysight” on page 4
Preventing Electrical Overstress Damage
Excessive RF power or voltage applied to the ports and connectors on the front
panel will damage the analyzer. Each port and front-panel connector has a
label (see photo below) that shows damage levels. These levels must not be
exceeded. Damage levels are also listed in the Specifications section of the
analyzer's Help system.
There are two ways to protect the PNA input ports from high RF power:
1.
Use power limiters at the ports. Information on limiters can be found at:
www.keysight.com/find/mta.
2.
Apply the recommendations in PNA Application Note 1408-10,
"Recommendations for Testing High-Power Amplifiers," found at:
http://literature.cdn.keysight.com/litweb/pdf/5989-1349EN.pdf
Description
Keysight Part Number
Grounding wrist strap
9300-1367
5-ft grounding cord for wrist strap
9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire
9300-0797