Installation Note E8356-90001 13
In This Guide
Installing Your Analyzer
Step 2. Meet Requirements
Before installing the PNA on a bench top or in an equipment rack, first make
sure that the electrical and environmental requirements of the location are met.
1. Ensure the available AC power source meets the following requirements:
2.
Verify that the AC power cable is not damaged, and that the power-source
outlet provides a protective earth contact.
This instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
Mains Supply Voltage
Frequency
Power
100/120 VAC
50/60/400 Hz
450 W MAX
220/240 VAC
50 /60 Hz
450W MAX
1) This instrument has an autoranging line voltage input.
Before switching on the instrument, be sure the supply
voltage is within the specified range and voltage
fluctuations do not exceed 10 percent of the nominal supply
voltage.
2) Condensing humidity operating conditions are not
allowed. Refer to “Environmental Information” on page 37.
3) The instruments can operate with mains supply voltage
fluctuations up to ± 10% of the nominal voltage.
Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord may
cause personal injury or product damage.