20 Installation Note E8356-90001
In This Guide
Installing Your Analyzer
-
Step 8. Run the Operator's Check
The operator's check should be performed when you first receive your analyzer,
and any time you wish to have confidence that the analyzer is working
properly. The operator's check does not verify performance to specifications,
but should give you a high degree of confidence that the instrument is
performing properly. The only equipment required is an open or short
calibration standard that mates with the front panel test ports. The open or
short calibration standard is NOT included with the PNA.
— On the A models: Press
System,
then
[Service]
softkey, then
[Verification]
softkey, then
[Operator's Check]
. Follow the prompts.
— On the B models: Press
System,
then
[Service]
softtab, then
[Verification]
,
then
[Operator's Check]
. Follow the prompts.
— For assistance, use the Help button displayed on the screen of the
Operator's Check.
If a "Log Error" is displayed during the verification stating that a
’filename.csv’ is already open and in use, verify that you only have a single
verification window open. This can be verified by mousing over the
verification icon on the toolbar as the bottom of the display. If there are
two verification windows open, close the window on the right most side,
which is the most recently opened window. And, resume the verification.