Installation Note E8356-90001 9
In This Guide
Operating Precautions
Visual Inspection
Do
—Inspect all connectors carefully
—Look for metal particles, scratches,
and dents
Do Not
—Use a damaged connector - ever
Connector Cleaning
Do
—Try compressed air first
—Use isopropyl alcohol (see warning
below)
—Clean connector threads
Do Not
—Use any abrasives
—Get liquid into plastic support
beads
Gaging Connectors
Do
—Clean and zero the gage before use
—Use the correct gage type
—Use correct end of calibration block
—Gage all connectors before first use
Do Not
—Use an out-of-specification
connector
Making Connections
Do
—Align connectors carefully
—Make preliminary connection
contact lightly
—Turn only the connector nut
—Use a torque wrench for final
connection
Do Not
—Apply bending force to connection
—Over tighten preliminary
connection
—Twist or screw any connection
—Tighten past torque wrench "break"
point
Keysight strongly recommends that adapters be used as connector savers
to minimize the wear on the analyzer test port connectors. Use
instrument-grade quality adapters or better.
Cleaning connectors with alcohol shall only be done with the
instrument power cord removed and in a well-ventilated area. Allow
all residual alcohol moisture to evaporate, and the fumes to
dissipate prior to energizing the instrument.