12
Keysight PD1000A Test Fixture Operating Guide
General Test Setup
General Test Setup
Figure 1 below shows the general test setup for the three PD1000A Test Fixtures
(the Surface Mount Device (SMD) fixture is shown in the photo). Note the two
PD1000A Bias T networks attached to the test fixture, the E5080A ENA Network
Analyzer behind the test fixture, and the B2902A Precision SMU (on top of the
E2080A ENA) providing bias voltage to the two Bias T networks.
Figure 1
General Test Setup for the PD1000A S-Parameter Test System
B2902A SMU
E5080A ENA
Device Test Fixture
Bias T
Bias T
Network Analyzer
B2902A Precision
Source/Measure Unit
for the Gate
for the Drain or Collector
(Surface Mount Device Test Fixture shown)
Do not excessively bend, flex, or stretch the SMA cables. Check the
SMA cables regularly, and replace them if they are damaged in any
way. It is possible for the outside shielding of the cables to break.