Keysight PD1000A Power Device Measurement System
23
Use Proper SMA Connector Care and Connection Techniques
3
Hold the torque wrench lightly, at the end of the handle only (beyond the
groove). See the following figure.
4
Apply force downward to the wrench handle. This applies torque to the
connection through the wrench. Do not hold the wrench so tightly that you
push the handle straight down along its length rather than pivoting it,
otherwise you apply an unknown amount of torque.
5
Tighten the connection just to the torque wrench break point. The wrench
handle gives way at its internal pivot point. Do not tighten the connection
further.
Separating Connections
To avoid lateral (bending) force on the connector mating plane surfaces, always
support the devices and connections.
1
Use an open-end wrench or spanner wrench to prevent the device body from
turning.
2
Use another open-end wrench or the torque wrench to loosen the connector
nut.
3
Complete the separation by hand, turning only the connector nut.
4
Pull the connectors straight apart without twisting, rocking, or bending.
Figure 9
Using the Torque Wrench
Torquing Direction
Stop when handle begins to yield
You don’t have to fully break the handle of the torque wrench to
reach the specified torque; doing so can cause the handle to kick
back and loosen the connection. Any give at all in the handle is
sufficient torque.
Do
not
turn the device body. Only turn the connector nut. Damage
to the center conductor can occur if the device body is twisted.