![Keysight Technologies L4490A Скачать руководство пользователя страница 98](http://html1.mh-extra.com/html/keysight-technologies/l4490a/l4490a_user-manual_1962268098.webp)
98
Keysight L4490A/L4491A User’s Guide
Introduction to Programming
System Verification
Running the Verification Sequence
The verification sequence is executed from the L4490A/L4491A Web interface.
As mentioned in Chapter 3, the L4490A/L4491A must be connected to the PC
through one of the LAN configurations shown in Figure 2-1. The Web interface is
not available from the GPIB interface. To use the Verification Board:
1
Turn off the L4490A/L4491A.
2
Install the verification board in bank 1 (or the desired bank) of the 34945EXT.
Securely fasten the verification board within the bank using the terminal
screws.
3
Turn on the L4490A/L4491A and open the web interface. Select the ‘Browser
Web Control’ page. See “Getting Started with the L4490A/L4491A
Web-Enabled Interface” for instructions on opening the Web page.
4
Select ‘Allow Full Control’ and in the ‘Browser Web Control’ window, locate
the bank and Y1156A programming diagram. An example of the diagram is
shown in Figure 4-8.
5
From the bank configuration window, select ‘
Preset Bank
’. It may be necessary
to turn off the channel drive (module configuration window) before ‘
Preset
Bank
’ is active.
6
From the remote module configuration window, set the
Channel Drive
source
to “Internal” or “External” depending on the 34945EXT supply and select
‘
Apply
’. The Y1156A is designed to use +24V supply.
7
Click the “Create Verification Sequence” button. This button defines the
verification sequence for the bank and opens the “Sequences...” window. With
the defined sequence highlighted, select ‘Execute Sequence’ to start the
sequence. The sequence will pass through the channel list three times. If the
signal path to the bank and each channel is complete, the LEDs located on
the board will illuminate as each channel in the list is closed.
8
Once the sequence completes, set the
Channel Drive
source to “Off”. Move the
Y1156A to the next bank. Click the “Update View” button located in the
upper-right corner of the ‘Browser Web Control’ window to refresh the Web
page with the new location of the Y1156A. Repeat steps 5-7 to test the signal
path to the channels in the next bank.
Содержание L4490A
Страница 1: ...Keysight L4490A L4491A RF Switch Platform User Guide...
Страница 3: ......
Страница 6: ...vi...
Страница 64: ...64 Keysight L4490A L4491A User s Guide Hardware Configuration Keysight Switch Attenuator Configuration...
Страница 126: ...126 Keysight L4490A L4491A User s Guide Introduction to Programming Power On RST and SYSTem CPON States...
Страница 154: ...12 Keysight L4490A L4491A User s Guide L4490A 91A Service Information...
Страница 158: ...16 Keysight L4490A L4491A User s Guide Index...
Страница 159: ...17...