![Keysight Technologies L4490A Скачать руководство пользователя страница 92](http://html1.mh-extra.com/html/keysight-technologies/l4490a/l4490a_user-manual_1962268092.webp)
92
Keysight L4490A/L4491A User’s Guide
Introduction to Programming
Getting Started with the L4490A/L4491A Web-Enabled Interface
This feature is limited to external channel drive sources (power supplies) with
the capability to drive multiple channels.
-
boot/reset channel drive - this parameter specifies the channel drive source
setting following a re-boot (power-up) or reset of the instrument.
Switch/Attenuator Channel Configuration Window
The switch/attenuator channel configuration window is opened by right-clicking
near the ‘IsOpen’ or ‘IsClosed’ labels on the ‘programming diagram’.
Figure 4-4. The Switch/Attenuator Channel Configuration Window.
The configuration window applies to the individual channels on the distribution
board. The parameters of this window include:
-
operation mode - enables channel pairing for dual-coil devices.
-
channel state after boot or Reset - specifies the channel state (opened, closed)
following power-on or reset (assumes boot/reset channel drive state is set).
-
activation type - specifies pulsed or continuous drive current.
-
pulse width - specifies the pulse width of the drive current.
-
recovery time - specifies the period between activation pulses
Содержание L4490A
Страница 1: ...Keysight L4490A L4491A RF Switch Platform User Guide...
Страница 3: ......
Страница 6: ...vi...
Страница 64: ...64 Keysight L4490A L4491A User s Guide Hardware Configuration Keysight Switch Attenuator Configuration...
Страница 126: ...126 Keysight L4490A L4491A User s Guide Introduction to Programming Power On RST and SYSTem CPON States...
Страница 154: ...12 Keysight L4490A L4491A User s Guide L4490A 91A Service Information...
Страница 158: ...16 Keysight L4490A L4491A User s Guide Index...
Страница 159: ...17...