6.4.3 Determining density of the test liquid
1. Fill test liquid into the glass beaker.
Make sure that the balance is in the mode for density determination of liquids (see
chap. 6.2.1).
If weighing balance does not show Zero, press
TARE
.
2. Place sinker in the upper sample dish, see fig. 5, chap. 6.1.2.
The weight of the sinker in air will be displayed.
3. Wait for stability display ( ), then press the
CAL
key.
4. Place sinker in the lower sifting bowl, see fig. 6, chap. 6.1.2..
The weighing balance will first determine then display the fluid’s density.
5. If you connect an optional printer you can print the result.
For further measurements
Clean and dry container and sinker carefully.
Re-attach the sinker
Press the
CAL
key
Start with step 2
To avoid corrosion, don't leave the density set immersed in liquid for a long time.