YDB-03-BA-e-2114
55
Special function keys:
Density determination of solids, see chap. 5.6.2
Density determination of liquids see chap. 6.6.2
Start measurement
Special info boxes:
Sequence of
operations
Selected type of density determination
(Method “solids“ or “liquids“)
Weighing
process 1
Weighing of sample in air
Weighing
process 2
Weighing of sample in liquid
Reference
liquid
Auxiliary liquid (distilled) water, ethanol or fluid of your choice of known
density.
Temperatur
e
Temperature of auxiliary liquid.
Density of
For density determination of solids:
Density of auxiliary liquid (for water or ethanol determined
automatically from the integrated density tables and then displayed)
For density determination of liquids:
Volume of plummet
5.7.1 Invoke mode for density determination of solid material and enter
parameters of the auxiliary liquid
To select method “solids“, press the
function key