5. Place the solid body under the lower strainer pan.
For that purpose, remove the dipper basket and after you have submerged it
again, place the sample under the strainer pan, trying to avoid formation of air
bubbles.
Or, whenever possible, use tweezers etc. to place the sample directly under the
strainer pan.
Fig. 4: Weighing in the auxiliary liquid
The sample weight in the auxiliary liquid will be displayed.
6. Wait until the stabilisation indicator is displayed (
), then press OK to take over
the weighed value. The density of the solid will be determined by the balance
taking into account aerodynamic displacement, and then displayed.
7. The result can be printed after an optional printer has been connected.
8. Remove the sample. To perform consecutive measurements, press OK and start
the procedure from step 3.
To prevent corrosion-related damage to the dipper basket, do not leave it
submerged in the liquid for an extended period of time.