![KERN KERN YDB-03 Скачать руководство пользователя страница 83](http://html.mh-extra.com/html/kern/kern-ydb-03/kern-ydb-03_operating-manual_1956872083.webp)
YDB-03-BA-e-2114
83
2. HOLD function <SG.HOLD>
Data-HOLD function can be activated both for the determination of density of solids
and liquids.
The displayed density value often fluctuates, and its reading may be difficult. When
the function is active, the first result value is shown in the display until it is cancelled
with the
UNIT
button.
Go to the menu:
In the weighing mode, press the
MENU
button twice.
Press the navigation buttons (
) until
“APL.FUNC” appears.
Confirm by pressing PRINT.
Press the navigation buttons (
) until
“SG” appears.
Confirm by pressing
TARE
. The display
shows “SET” and the current setting.
Press the navigation buttons (
) until
“SG.HOLD” appears.
Confirm by pressing
TARE
.
By pressing
TARE
, select between the
settings “OFF” and “ON”. The current
setting is marked by the stabilization
indicator.
Stabilization indicator
“SG.HOLD” setting
OFF
OFF
ON
ON
Содержание KERN YDB-03
Страница 7: ...YDB 03 BA e 2114 7 Adapter 5 items Allen wrench screw ...
Страница 8: ...8 YDB 03 BA e 2114 2 Dimension mm Frame ...
Страница 9: ...YDB 03 BA e 2114 9 Universal immersion basket for floating and descending solid matter Sample dish ...
Страница 10: ...10 YDB 03 BA e 2114 Platform Beaker ...
Страница 90: ...90 YDB 03 BA e 2114 Back to menu weighing mode press ON OFF key repeatedly ...