6.9.3 Determination of the examined liquid density
1. Fill the beaker with the examined liquid. Attach the dipper basket. It should not
touch the beaker. Press
TARE
to reset the balance.
2. Ensure the balance is in the <
Liquid density> mode (see chapter 6.8.2).
Reset the balance as required.
3. Place the sinker on the upper sample pan.
The sinker weight in the air will be displayed.
4. Wait until the stabilisation indicator is displayed (
), then press OK to take over
the weighed value.
5. Place the sinker on the lower strainer pan.
To do it, remove the dipper basket from the stand. When submerging in the liquid
again, no extra air bubbles should be created. It is best to apply the sample using
tweezers or place it directly on the strainer pan.