72
YDB-03-BA-e-2114
Method 2:
1. Place a beaker filled with the auxiliary liquid in the platform centre. It should be
filled to ca. 3/4
th
of its capacity. It should not touch the stand.
Attach the dipper basket. It should not touch the beaker. Reset the balance.
2. Ensure the balance is in the <
Solid specific gravity> mode (see chapter 5.7.2).
3. Place the solid body on the upper sample pan.
Fig. 3: Weighing in the air
The sample weight in the air will be displayed.
4. Wait until the stabilisation indicator is displayed (
), then press OK to take over
the weighed value.
Содержание KERN YDB-03
Страница 7: ...YDB 03 BA e 2114 7 Adapter 5 items Allen wrench screw ...
Страница 8: ...8 YDB 03 BA e 2114 2 Dimension mm Frame ...
Страница 9: ...YDB 03 BA e 2114 9 Universal immersion basket for floating and descending solid matter Sample dish ...
Страница 10: ...10 YDB 03 BA e 2114 Platform Beaker ...
Страница 90: ...90 YDB 03 BA e 2114 Back to menu weighing mode press ON OFF key repeatedly ...