FIS-BA-e-0512
31
8.2
Input of nominal value and tolerance limits, function start
The
F
key is assigned to a tolerance check input function.
Input function selection is carried out as described in chapter 6.4.
8.2.1 Weighed nominal value and tolerances
The
F
key is assigned to
Note: See chapters 6.2 – 6.4 for assignment of the function key
A nominal weight of 2.000 kg is
on the balance
F
Start the tolerance check
function.
The nominal weight is weighed
on start-up of the function.
Weigh the lower Tu tolerance.
Apply a Tu weight of 1.800 kg
The weight value is indicated
F
Start weighing
The value for Tu has been
adopted.
Weigh the upper To tolerance
A To weight of 2.100 kg has
been applied
F
Start weighing
The value for To has been
adopted
The tolerance check is started after the To has been measured
The green pilot lamp indicates the result of the weight evaluation. Further weights can now
be checked. Nominal value and tolerance limits are maintained until the function is
switched off using the
F
key.