FIS-BA-e-0512
53
14 Miscellaneous
14.1
Disturbances, causes, remedial action
Continuous automatic
check and test routines are carried out.
Status and error
messages
are the
result
of these test routines.
Soft Error
Activate the
F
key. This permits operations to
be continued on the weighing terminal
Hard Error
The error cannot be corrected.
Activating the
ON
OFF
key causes a function test to be
started
Disturbances
Causes
Remedial Action
No power supply or plug
not connected
Connect the direct plug-in transformer
unit
Pre-load missing or load
plate jamming.
Apply the pre-load.
Ensure that load plate has free play.
Weighing range exceeded. Unload the scale.
Vibrations at place of
installation
Material to be weighed in
contact with other parts
Weighing material unstable
Eliminate disturbances at the place of
installation.
Adapt parameter settings.
Material to be weighed not
positioned properly on the
scale. In contact with other
parts.
Position the material to be weighed
properly.
Position it in such a manner as to
prevent it from coming into contact with
other parts.
Error in weighing terminal.
Error in load carrier
Activate the
ON
OFF
key;
Inform KERN quoting the error
number.
Activate the
ON
OFF
key to remedy
momentary disturbances.