FIS-BA-e-0512
54
The most important error numbers
ADW error numbers
Err 20951
RAM error
Err 20952
ROM error
Err 20100
Weighing cell parameter too small for verifiable balances
(acknowledge using the tare cancel key)
Err 20101
Weighing cell zero point too small
Err 20102
Weighing cell zero point too large
Err 20800
Faulty data entered ( divided by zero )
Err 20851
ADW timeout, no ADW interrupt
Err 20900
Error on pre-load adaptation
EEPROM
error numbers
Err 21049
Calibration attempt without EEPROM plug-in bridge
Err 21050
EEPROM writing error, e.g. calibration bridge not plugged in
Err 21051
Checksum error verifiable data RAM
Err 21052
Checksum error EEPROM block 1
Err 21053
Checksum error EEPROM block 2
Err 21054
Checksum error EEPROM block 3
Err 21055
Checksum error EEPROM block 4
Err 21056
Checksum error consecutive number