5 Calibration
Information
Weight symbol
flashing:
Calibration switch (hardware seal) is
not
in secured condition.
Calibration switch (=equalising plug
)
Traditionally, backup of the data relevant to calibration is carried out by means of
hardware backup (equalising plug on the circuit board in the terminal).
Equalising plug
on the circuit board in the terminal
Status
Plug not inserted
Calibration data not saved
Plug inserted
Calibration data saved
The calibration switch (equalising plug) can be seen through the right-hand window
on the terminal. This switch is secured by means of a sticker (stamp) when in
calibrated condition.
Balance calibration is not valid without the valid sticker.
The four screws on the rear side of the terminal must be opened in order to reach
the calibration switch (equalising plug).
Caution:
Please remove the mains plug before opening the terminal.
The GEO value can only be changed or the balance calibrated if the calibration
switch (equalising plug) has been pulled out.