Using both sides of the test fixture
In several of the interconnection diagrams, it is noted that you can
make equivalent connections on the other side of the test fixture. For
example, if you have an ongoing test using the Series 2600 connected
to the high-current side, you could set up a one-time test using the
equivalent terminals on the high-voltage side.
Note that this option is not available for test setups when you are
using a Model 2651 or Model 2657.
Do not simultaneously connect terminals in the high current and
high voltage areas of the fixture to the same terminals in the
center area (the area bordered in blue). Simultaneous
connections can result in electric shock (see previous graphic).
Содержание 8010
Страница 1: ...Model 8010 High Power Device Test Fixture Interconnection Reference Guide ...
Страница 13: ...External wiring diagram current voltage measurement connections ...
Страница 14: ...Two terminal axial lead DUT with a Model 2651 connected local sense lid open view of device test boards ...
Страница 16: ...Two terminal axial lead DUT with a Model 2657 connected local sense ...
Страница 20: ...Two terminal axial lead DUT with a Model 4200 SMU connected local sense ...
Страница 22: ...This page intentionally left blank ...
Страница 24: ...Three terminal DUT with a Model 2657 and a Series 2600 connected local sense ...
Страница 25: ...Three terminal DUT with a Model 2657 and a Series 2600 connected remote sense ...
Страница 28: ...Three terminal DUT with a 4200 and 2657 connected local sense ...
Страница 29: ...Three terminal DUT with a 4200 and 2651 connected remote sense ...
Страница 31: ...This page intentionally left blank ...
Страница 32: ......
Страница 33: ......
Страница 35: ...CVU 200 KIT 2 terminal DUT local sense ...
Страница 36: ......