Assumptions for connection diagrams
When you use the connection diagrams in this Interconnection
Reference Guide, be aware of the following assumptions.
SourceMeter
®
Instrument information
You cannot use a Model 2651 and Model 2657 simultaneously to test
a single device.
You can use a Model 2651 or Model 2657 simultaneously with a
Series 2600 instrument. When a Series 2600 is referenced in the
Model 8010 documentation, you can use any of the following
SourceMeter Instruments:
Model 2611
Model 2612
Model 2635
Model 2636
Using remote sense
If you are using remote sense connections, you must enable remote
sense on the instrument.
Types of devices
All three-terminal devices under test (DUTs) are assumed to be
MOSFETs. Therefore, the labels G-D-S on the connection diagrams
relate to the Gate, Drain, and Source of a MOSFET device.
If you are testing an insulated gate bipolar transistor (IGBT), make
the following substitutions:
G: Connect to the gate
D: Connect to the collector
S: Connect to the emitter
If you are testing a bipolar junction transistor, make the following
substitutions:
G: Connect to the base
D: Connect to the collector
S: Connect to the emitter
Note: Remove the 4200-PA Remote Preamplifier when the
4200-SCS is connected to the Model 8010 or to the CVU-3K-KIT
bias tees.
Содержание 8010
Страница 1: ...Model 8010 High Power Device Test Fixture Interconnection Reference Guide ...
Страница 13: ...External wiring diagram current voltage measurement connections ...
Страница 14: ...Two terminal axial lead DUT with a Model 2651 connected local sense lid open view of device test boards ...
Страница 16: ...Two terminal axial lead DUT with a Model 2657 connected local sense ...
Страница 20: ...Two terminal axial lead DUT with a Model 4200 SMU connected local sense ...
Страница 22: ...This page intentionally left blank ...
Страница 24: ...Three terminal DUT with a Model 2657 and a Series 2600 connected local sense ...
Страница 25: ...Three terminal DUT with a Model 2657 and a Series 2600 connected remote sense ...
Страница 28: ...Three terminal DUT with a 4200 and 2657 connected local sense ...
Страница 29: ...Three terminal DUT with a 4200 and 2651 connected remote sense ...
Страница 31: ...This page intentionally left blank ...
Страница 32: ......
Страница 33: ......
Страница 35: ...CVU 200 KIT 2 terminal DUT local sense ...
Страница 36: ......