Connect the test fixture external wiring (current-voltage)
Important test system safety information
This product is sold as a stand-alone test fixture that may become
part of a system that could contain hazardous voltages and energy
sources. It is the responsibility of the test system designer, integrator,
installer, maintenance personnel, and service personnel to make sure
that the system is safe during use and that it is operating properly.
It is important that you consider the following factors in your system
design and use:
The international safety standard UL 61010-1: 2004 defines
voltages as hazardous if they exceed 30 V RMS and 42.4 V peak,
or 60 V DC for equipment rated for dry locations. Keithley
Instruments products are only rated for dry locations.
Read and comply with the specifications of all instruments in the
system. The overall allowed signal levels may be constrained by
the lowest rated instrument in the system. For example, if you are
using a 500 V power supply with a 300 V DC rated switch, the
maximum allowed voltage in the system is 300 V DC.
Make sure any test fixture connected to the system protects the
operator from contact with hazardous voltages, hot surfaces, and
sharp objects. Use shields, barriers, insulation, and safety
interlocks to accomplish this.
Provide training to all system users so that they understand all
potential hazards and know how to protect themselves from
injury.
Never connect the guard terminal from any instrument to the LO
terminal of any instrument in the Model 8010 or to the chassis.
Connecting guard to LO can disable the high voltage protection
that is installed across the Model 4200 or Model 2600
SourceMeter Instrument connections. This may result in
hazardous live voltages being present at the HI, SHI, SLO, or LO
terminals.
Содержание 8010
Страница 1: ...Model 8010 High Power Device Test Fixture Interconnection Reference Guide ...
Страница 13: ...External wiring diagram current voltage measurement connections ...
Страница 14: ...Two terminal axial lead DUT with a Model 2651 connected local sense lid open view of device test boards ...
Страница 16: ...Two terminal axial lead DUT with a Model 2657 connected local sense ...
Страница 20: ...Two terminal axial lead DUT with a Model 4200 SMU connected local sense ...
Страница 22: ...This page intentionally left blank ...
Страница 24: ...Three terminal DUT with a Model 2657 and a Series 2600 connected local sense ...
Страница 25: ...Three terminal DUT with a Model 2657 and a Series 2600 connected remote sense ...
Страница 28: ...Three terminal DUT with a 4200 and 2657 connected local sense ...
Страница 29: ...Three terminal DUT with a 4200 and 2651 connected remote sense ...
Страница 31: ...This page intentionally left blank ...
Страница 32: ......
Страница 33: ......
Страница 35: ...CVU 200 KIT 2 terminal DUT local sense ...
Страница 36: ......