Insulating plug use
The Model 8010 comes with insulating plugs that can be used when
testing two-terminal devices.
The three-pin socket inherently shorts the force and sense pins
together. When you are using the axial terminal posts in 4-wire mode,
this can cause measurement errors. You should insert the insulating
plug into the three-pin socket when testing axial-lead devices in
4-wire sense mode where a short between force and sense will cause
measurement errors.
Do not use the insulating plug when testing higher resistance devices
(greater than 1 M
Ω
). Using the plug for these tests may cause leakage
measurement errors.
Testing two-terminal TO-247 devices
The two-terminal interconnection diagrams that are shown in this
document are intended for axial-lead devices. However, you can
adapt the setups so that you can test two-terminal TO-247 devices.
To test TO-247 devices, you need to:
1. Remove the insulating plug (if used) from the three-pin socket.
2. Verify that the four axial device terminal posts are not connected
to anything.
3. Place the two-terminal TO-247 device into the three-pin socket.
Содержание 8010
Страница 1: ...Model 8010 High Power Device Test Fixture Interconnection Reference Guide ...
Страница 13: ...External wiring diagram current voltage measurement connections ...
Страница 14: ...Two terminal axial lead DUT with a Model 2651 connected local sense lid open view of device test boards ...
Страница 16: ...Two terminal axial lead DUT with a Model 2657 connected local sense ...
Страница 20: ...Two terminal axial lead DUT with a Model 4200 SMU connected local sense ...
Страница 22: ...This page intentionally left blank ...
Страница 24: ...Three terminal DUT with a Model 2657 and a Series 2600 connected local sense ...
Страница 25: ...Three terminal DUT with a Model 2657 and a Series 2600 connected remote sense ...
Страница 28: ...Three terminal DUT with a 4200 and 2657 connected local sense ...
Страница 29: ...Three terminal DUT with a 4200 and 2651 connected remote sense ...
Страница 31: ...This page intentionally left blank ...
Страница 32: ......
Страница 33: ......
Страница 35: ...CVU 200 KIT 2 terminal DUT local sense ...
Страница 36: ......