Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components
in mains circuits — including the power transformer, test leads, and input jacks
— must be purchased from Keithley Instruments. Standard fuses with
applicable national safety approvals may be used if the rating and type are the
same. The detachable mains power cord provided with the instrument may only
be replaced with a similarly rated power cord. Other components that are not
safety-related may be purchased from other suppliers as long as they are
equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and
functionality of the product). If you are unsure about the applicability of a
replacement component, call a Keithley Instruments office for information.
Unless otherwise noted in product-specific literature, Keithley instruments are
designed to operate indoors only, in the following environment: Altitude at or
below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and
pollution degree 1 or 2.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean
the exterior of the instrument only. Do not apply cleaner directly to the
instrument or allow liquids to enter or spill on the instrument. Products that
consist of a circuit board with no case or chassis (e.g., a data acquisition board
for installation into a computer) should never require cleaning if handled
according to instructions. If the board becomes contaminated and operation is
affected, the board should be returned to the factory for proper
cleaning/servicing.
Safety precaution revision as of March 2016.
Power and environmental characteristics
For indoor use only.
Maximum
signal voltage
(signal or
guard to any
signal)
Three-lug high-voltage triaxial connector: 3280 V
Three-lug standard triaxial connector: 210 V
Eight-pin screw terminal connector: 40 V
Two-pin high-current screw terminal connector: 40 V
Maximum
signal current
Three-lug high-voltage triaxial connector: 120 mA DC
Three-lug standard triaxial connector: 1.5 A DC
Eight-pin screw terminal connector: 1 A DC
Two-pin high-current screw terminal connector: 15 A DC,
50 A pulsed with one SMU; 100 with two SMUs
Maximum
combined DC
current
15 A DC
Maximum
pulse current
100 A at 1 % duty cycle for Model 2651 High Power
SourceMeter Instrument path
10 A at 1 % duty cycle for Series 2600 and Model 4200 paths
Altitude
Maximum 2000 m above sea level
Operating
0 °C to 50 °C, 70 % relative humidity up to 35 °C
Derate 3 % relative humidity/°C, 35 °C to 50 °C
Storage
-25 °C to 65 °C
Safety
Listed to UL61010-1:2004
Conforms to European Union Low Voltage Directive
Содержание 8010
Страница 1: ...Model 8010 High Power Device Test Fixture Interconnection Reference Guide ...
Страница 13: ...External wiring diagram current voltage measurement connections ...
Страница 14: ...Two terminal axial lead DUT with a Model 2651 connected local sense lid open view of device test boards ...
Страница 16: ...Two terminal axial lead DUT with a Model 2657 connected local sense ...
Страница 20: ...Two terminal axial lead DUT with a Model 4200 SMU connected local sense ...
Страница 22: ...This page intentionally left blank ...
Страница 24: ...Three terminal DUT with a Model 2657 and a Series 2600 connected local sense ...
Страница 25: ...Three terminal DUT with a Model 2657 and a Series 2600 connected remote sense ...
Страница 28: ...Three terminal DUT with a 4200 and 2657 connected local sense ...
Страница 29: ...Three terminal DUT with a 4200 and 2651 connected remote sense ...
Страница 31: ...This page intentionally left blank ...
Страница 32: ......
Страница 33: ......
Страница 35: ...CVU 200 KIT 2 terminal DUT local sense ...
Страница 36: ......