Model 6487 Reference Manual
Measurements and Sourcing Voltage
3-17
Step 2. Connect the load to the source output
Basic connections for sourcing voltage are shown in
.
t^okfkd
A safety shield is advisable whenever measurements are being made
with voltages over 30V DC. Connections for the safety shield are
shown in
. The metal safety shield must completely surround
the noise shield or floating test circuit and it must be connected to
safety earth ground using #18 AWG or larger wire.
Step 3. Turn on voltage source
Press the OPER key to turn on the voltage source output. The VOLTAGE SOURCE
OPERATE indicator will turn on.
`^rqflk
Do not connect external sources to the 6487 voltage source. External
sources may damage the 6487 voltage source.
Figure 3-4
Connections for sourcing voltage
Me
t
al
S
a
f
e
t
y
S
hiel
d
S
a
f
e
t
y
Ea
rt
h
Gr
oun
d
Model 64
8
7
V-
S
OURCE
OUTPUT
Loa
d
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
INTERLOCK
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI
!
!
FUSE
LINE
400mAT
(SB)
100 VAC
120 VAC
200mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
Содержание 6487
Страница 185: ...11 CommonCommands Common commands This section lists and describes the common commands...
Страница 190: ...12 SCPISignalOriented MeasurementCommands...
Страница 254: ...A Specifications...
Страница 256: ...B StatusandErrorMessages...
Страница 264: ...C DDCEmulationCommands...