7-8
Triggering
Model 6487 Reference Manual
Trigger model configuration — front panel
klqb
See “SCPI Programming” (
) for the SCPI commands to configure the
trigger model over the bus.
Press CONFIG and then TRIG to configure both the TRIG and ARM layers of the trigger
model.
klqb
When done configuring the trigger level, press ENTER to confirm value and then
use the EXIT key to save changes and leave trigger model configuration.
Once in trigger model configuration mode, use the RANGE keys to display either the
TRIGGER layer or the ARM layer menus. Press ENTER to select the desired menu. Then
use the RANGE keys to display menu items. Use the cursor keys to key in values. A menu
item or value is selected by pressing ENTER. Use the EXIT key to exit from the menu
(saving all changes made).
Table 7-2
Trigger model menu structure
Menu
Description
CONFIG
Path to TRIG and ARM menus. Access by pressing
CONFIG and then TRIG.
-
TRIG
Path to TRIG layer sub-menus.
- -
COUNT
- - - INF
- - - FIN
Set TRIG measure count.
Specify an INFinite measure count.
Specify a FINite measure count: 1–2048.
- -
DELAY
- - - MAN
- - - AUTO
Set trigger delay.
Specify trigger delay: 0–999.9998 sec.
Enable auto delay.
- -
TRIG-OUT
- - - LINE
- - - - TLINK
- - - EVENTS
- - - - VMC
Configure output triggers.
Select the output trigger link line.
Set trigger link value: 1–6.
Select VMC (voltmeter complete) output trigger.
Enable/disable VMC (on or off).
- -
TRIG-IN
- - - SOURCE
- - - - IMM
- - - - TLINK
- - - - - TLINK
- - - - - EVENT
Path to control source.
Set the TRIG-IN control source.
Set control source to IMMediate.
Set control source to TLINK.
Set TLINK value: 1–6.
Enable (set to ONCE) or disable (set to NEVER).
Содержание 6487
Страница 185: ...11 CommonCommands Common commands This section lists and describes the common commands...
Страница 190: ...12 SCPISignalOriented MeasurementCommands...
Страница 254: ...A Specifications...
Страница 256: ...B StatusandErrorMessages...
Страница 264: ...C DDCEmulationCommands...