Programming example
The following command sequence performs a sweep from 1V to 10V in 1V increments
and recalls all readings:
*RST
' Return 6487 to RST defaults.
SOUR:VOLT:SWE:STAR 1
' Start voltage = 1V.
SOUR:VOLT:SWE:STOP 10
' Stop voltage = 10V.
SOUR:VOLT:SWE:STEP 1
' Step voltage = 1V.
SOUR:VOLT:SWE:DEL 0.1
' 0.1s delay.
ARM:COUN INF
' Reset arm count.
FORM:ELEM READ,VSO
' Select reading, voltage source data.
SOUR:VOLT:SWE:INIT
' Arm sweep, put source in operate.
SYST:ZCH OFF
' Turn off zero check.
INIT
' Trigger sweep.
TRAC:DATA?
' Request all stored sweep readings.