Model 6487 Reference Manual
Remote Calibration
F-5
Table F-3
Voltage source calibration commands by range
Range
Commands*
10V
:SOUR1:VOLT:RANG 10
:CAL:PROT:VSO:NFSC
:CAL:PROT:VSO:NFSV <DMM_Reading>
:CAL:PROT:VSO:ZERO
:CAL:PROT:VSO:ZVAL <DMM_Reading>
:CAL:PROT:VSO:PFSC
:CAL:PROT:VSO:PFSV <DMM_Reading>
50V
:SOUR1:VOLT:RANG 50
:CAL:PROT:VSO:NFSC
:CAL:PROT:VSO:NFSV <DMM_Reading>
:CAL:PROT:VSO:ZERO
:CAL:PROT:VSO:ZVAL <DMM_Reading>
:CAL:PROT:VSO:PFSC
:CAL:PROT:VSO:PFSV <DMM_Reading>
500V
:SOUR1:VOLT:RANG 500
:CAL:PROT:VSO:NFSC
:CAL:PROT:VSO:NFSV <DMM_Reading>
:CAL:PROT:VSO:ZERO
:CAL:PROT:VSO:ZVAL <DMM_Reading>
:CAL:PROT:VSO:PFSC
:CAL:PROT:VSO:PFSV <DMM_Reading>
* Include <DMM_Reading> from output voltage generated by previous step.
For example, :CAL:PROT:VSO:NFSV <DMM_Reading> value is generated by
:CAL:PROT:VSO:NFSC command.