![Keithley 6487 Скачать руководство пользователя страница 245](http://html1.mh-extra.com/html/keithley/6487/6487_reference-manual_4064220245.webp)
Model 6487 Reference Manual
Calibration
16-11
Figure 16-2
Connections for 2nA to 2µA range calibration
Table 16-6
2nA to 2µA range calibration summary
Model 6487 range
Calibrator voltages
Standard resistors
1
Calibration currents
2
2nA
0V
+2.000000V
-2.000000V
1G
Ω
1G
Ω
1G
Ω
0nA
+2nA
-2nA
20nA
0V
+2.000000V
-2.000000V
100M
Ω
100M
Ω
100M
Ω
0nA
+20nA
-20nA
2
μ
A
0V
+200.0000V
-200.0000V
100M
Ω
100M
Ω
100M
Ω
0
μ
A
+2μ
A
-2
μ
A
1.
1
Nominal resistance values.
2.
2
Nominal currents. Calculate actual currents from calibrator voltage and actual standard resistor value: I = V/R. Calibrate zero,
positive full-scale, and negative full-scale for each range.
I
N
PUT
T
r
iax Cable
Connec
t
Calib
r
a
t
o
r
t
o
100M
Ω
o
r
1
G
Ω
Re
s
i
st
o
r
Link
S
hiel
d
an
d
Cha
ss
i
s
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI INTERLOCK
!
!
FUSE
LINE
400mAT
(SB)
100 VAC
120 VAC
200mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
Model 64
8
7 Picoamme
t
e
r
DC Vol
t
age Calib
r
a
t
o
r
B
N
C-
t
o-
d
ual
Banana Plu
g
A
d
ap
t
e
r
100
G
Ω
100M
Ω
10
G
Ω
1
G
Ω
1nF
100nF
OUTPUT
Connec
t
Cable
S
hiel
d
t
o
Ou
t
pu
t
LO
Low-noi
s
e
Coax Cable
N
o
t
e:
Model 5156 Calib
r
a
t
ion S
t
anda
r
d
!
Содержание 6487
Страница 185: ...11 CommonCommands Common commands This section lists and describes the common commands...
Страница 190: ...12 SCPISignalOriented MeasurementCommands...
Страница 254: ...A Specifications...
Страница 256: ...B StatusandErrorMessages...
Страница 264: ...C DDCEmulationCommands...